Page 37 E
VERFLO
TM
SERVICE & TECHNICAL INFORMATION 1039055, VER. 06
SYMPTOM CAUSE VERIFICATION
CORRECTIVE
ACTION
•
Pressure relief
valve activated
•
Red LED (constant)
•
Low pressure
Cycle failure
•
Verify Valve Solenoid
Assembly operation
•
Verify Valve Solenoid
assembly wire
harness connection
with Main PCA
•
Verify voltage from
Main PCA to Valve
Solenoid Assembly
•
Replace Valve
Solenoid Assembly
•
Reconnect the Valve
Solenoid wiring
harness to the Main
PCA
•
Replace the Main PCA
•
Inaccurate oxygen
readings
Leaks
Main PCA failure
Compressor failure
•
Verify the hoses are
connected to the OPI
sensor.
•
Verify that there are
no leaks
•
Verify the output with
a calibrated oxygen
analyzer
•
Verify the
Compressor output
range.
•
Reconnect the
hoses
•
Reconnect the
tubing
•
Replace Main PCA
•
Replace the
Compressor