1 – Connect indication request for the LAN interface. The Connect Indication message is only
available on the LAN interface. As soon as the IntelliVue monitor has received a valid IP address from
the BOOTP server in the network, the monitor sends out the Connect Indication message of its LAN
interface.
2 – Association Request accepted. To establish a logical connection, the Computer Client sends the
Association Request message to the IntelliVue monitor. The IntelliVue monitor processes the
Association Request and sends an Association Result.
Type
The type of data export connection is displayed: Either LAN (IEEE802.3 10BaseT (10MBit/s) Local
Area Network) or COM (RS-232 or RS-232 via USB adapter interface or "RS-232 over LAN
adapter").
Name
The Associated Name of the patient monitor system or the Interface is displayed.
LAN: If the connected patient monitor uses a label it will be displayed.
COM: The respective COM Port No. together with the interface speed will be displayed, e.g.
COM1:115200 (COM1 with 115,200 speed)
COM/MAC
LAN: MAC Address of the IntelliVue patient monitor
COM: Assigned COM Interface number of the PC
IP
IP Address assigned to the LAN Interface of the IntelliVue patient monitor.