Diagnostic Software
EN 59DVDR615, MRV640 5.
Nucleus Name DS_SYS_DisplayFatalOn
Nucleus Number 1223
Description Turn on the display-fatal functionality which displays debug-information on the dis-
play when encountering a fatal error condition from which could not be recovered au-
tomatically
Technical - Issue the command to use the display-fatal functionality to the analogue board
- Wait for the result and return this to the user
Execution Time Less than 1 second.
User Input None
Error Number Description
122300 Turning on the display-fatal functionality succeeded
122301 Turning on the display-fatal functionality failed
122302 The returned errorcode from the analogue board is un-
known:
122303 No DS errCode known for analogue board error:
122304 There was no response from the analogue board.
Example DS:> 1223
122300:
Test OK @
Nucleus Name DS_SYS_DisplayFatalOff
Nucleus Number 1224
Description Turn off the display-fatal functionality which displays debug-information on the dis-
play when encountering a fatal error condition from which could not be recovered au-
tomatically
Technical - Issue the command to stop using the display-fatal functionality to the analogue
board
- Wait for the result and return this to the user
Execution Time Less than 1 second.
User Input None
Error Number Description
122400 Turning off the display-fatal functionality succeeded
122401 Turning off the display-fatal functionality failed
122402 The returned errorcode from the analogue board is un-
known:
122403 No DS errCode known for analogue board error:
122404 There was no response from the analogue board.
Example DS:> 1224
122400:
Test OK @
Nucleus Name DS_SYS_DisplayFatalGet
Nucleus Number 1225
Description Get the display-fatal flag of the recorder
Technical - Issue the command to get the status of the display-fatal functionality to the
analogue board
- Wait for the result and return this to the user
Execution Time Less than 1 second.
User Input None
Error Number Description
122500 Getting the display-fatal flag succeeded
122501 Getting the display-fatal flag failed
122502 The returned errorcode from the analogue board is un-
known:
122503 No DS errCode known for analogue board error:
122504 There was no response from the analogue board.
Example DS:> 1225
122500:
Test OK @
Nucleus Name DS_SYS_SettingsSet
Nucleus Number 1226
Description Programs the digital board settings into the boot EEPROM on the digital board.