Battery Pack Care
Leaving the 2020PROPlus for more than 3 months without charging the batteries
may result in the loss of recorded data and setup parameters. If the 2020PRO
Plus is
not used for long periods of time, recharge the battery for a few hours every 30 days
to avoid loss of data. See Battery Charging on page 39 for instructions on charging
the battery.
Please observe the following:
• For replacement battery pack use only Part No. MX700010.
• Do not dispose of the battery pack in a fire. The cell may explode.
• The battery pack is 24% cadmium (Cd) by weight. This battery pack must be
disposed of properly. Check with local codes for special disposal instructions.
• Do not open or mutilate the battery pack.
• Exercise care in handling battery packs in order not to short the terminals with
conducting materials such as rings, bracelets and keys. The battery or conductor
may overheat and cause burns.
• Charge the battery pack using the AC adapter provided with or identified for use
with this product only in accordance with the instructions and limitations
specified in this manual. For AC adapter use only Part No. MX350002 (North
America) or MX396013 (Europe).
Maintenance of the UV Lamp
Removing and Replacing the UV Lamp
NOTE: Do not remove or replace the UV lamp in a hazardous location.
To remove the UV lamp from your instrument:
1. Ensure the instrument is turned off.
WARNING
You must turn the instrument off before removing the lamp cover.
2. Remove the lamp housing cover.
3. Tilt the 2020PRO
Plus slightly and remove the UV lamp.