LUCAS
TM
2 Chest Compression System –
Instructions for Use
100901-00 Rev A, Valid from CO J2716 © 2014 Jolife AB
35
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
15
0 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.24
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
d 1.2 P=
d 1.2 P=
d 1.3 P=
100901-00 Rev B, Valid from CO J3166 © 2017 Jolife AB