Confirm cardiac arrest and start manual CPR with a minimum of interruptions until
LUCAS is applied and ready.
Always follow local and/or international guidelines for CPR when you use LUCAS.
ON/OFF ADJUST PAUSE ACTIVE
(continuous)
ACTIVE
(30:2)
QUICK REFERENCE GUIDE
LUCAS is manufactured by JOLIFE AB in Sweden and distributed worldwide by Physio-Control, Inc.
For further information, please visit our website at www.physio-control.com.
NOTE: This guide is not a complete Instructions for Use. Refer to the “Instructions for Use” for complete
directions for use, indications, contraindications, warnings, precautions, and potential adverse events.
LUCAS™ 2 Chest Compression System – Instructions for Use
100901-00 Rev A, Valid from CO J2716 © 2014 Jolife AB
Activate (A)
s Push ON/OFF for 1 second to
start self-test and power up
LUCAS
Back Plate (B)
s Pause manual CPR
s Carefully put Back Plate under
the patient, below armpits
s Resume manual CPR
Compressor (C)
s Pull release rings once; claw
locks open. Then let go of the
release rings
s Attach to Back Plate; listen
for “click”
s Pull up once to ensure
attachment
Position the
Suction Cup
s Center the Suction Cup over
the chest
s The lower edge of Suction Cup
should be immediately above
the end of the sternum
Push down the
Suction Cup
s Push the Suction Cup down
with two fingers (make sure
it is in the ADJUST mode)
s Pressure pad inside Suction
Cup should touch patient’s
chest. If the pad does not
touch or fit properly, continue
manual compressions
s Push PAUSE to lock Start
Position – then remove your
fingers from the Suction Cup
Start compressions
s Check for proper position.
Adjust if necessary
s Push ACTIVE (continuous)
or ACTIVE (30:2)
s LUCAS provides chest
compressions according to
guidelines.
Attach the
Stabilization Strap
s Attach the LUCAS Stabilization
Strap