28
EssentiA
+
Instruction Manual
Electromagnetic Compatibility Information
Table 3
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such
an environment.
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-4-3
Portable and mobile RF
communications equipment
should be used no closer to any
part of the device, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended seperation distance
80 MHz to 800 MHz
800 MHz to 2.7 Ghz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electro magnetic site survey, a
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
3V for 0.15-
80MHz; 6V
in ISM and
amateur radio
bands between
0.15-80MHz
3V for 0.15-
80MHz; 6V
in ISM and
amateur radio
bands between
0.15-80MHz
450MHz, 28V/m 450MHz, 28V/m
710MHz,745
MHZ,780MHz
9V/m
810MHz,870
MHZ,930MHz
28V/m
1720MHz,1845
MHZ,1970MHz
28V/m
2450MHz,
28V/m
5240MHz,5500
MHZ,5785MHz
9V/m
710MHz,745
MHZ,780MHz
9V/m
810MHz,870
MHZ,930MHz
28V/m
1720MHz,1845
MHZ,1970MHz
28V/m
2450MHz,
28V/m
5240MHz,5500
MHZ,5785MHz
9V/m
385MHz, 27V/m 385MHz, 27V/m
Electromagnetic environment
-guidance
Compliance
level
IEC 60601
test level
IMMUNITY test