Electromagnetic Compatibility Information
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such
an environment.
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electromagnetic environment
-guidance
Compliance
level
IEC 60601
test level
IMMUNITY test
± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±0.5kV, ±1kV
(differential
mode)
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m; 50Hz
or 60Hz
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
30 A/m; 50Hz
or 60Hz
± 8 kV
contact
±2 kV,±4 kV,
±8 kV,
±15 kV air
± 2 kV ,
100kHz, for
AC power port
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300 cycle
±0.5kV, ±1kV
(differential
mode)
.
± 2 kV ,
100kHz, for
AC power port
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.