SECTION 7 - MAINTENANCE INFORMATION
Page 7.5
pickering
ISOLATED MILLIVOLT THERMOCOUPLE SIMULATOR 41-761
Table 7.2 - Verication Test Results For +/-100mV Range (Accuracy: 0.1% ±15μV)
Serial Number:
Channel Number:
Date of Test:
Operators Name:
Test Equipment
Used:
Voltage (mV) Measured Value Maximum (mV) Minimum (mV)
Result (delete
as appropriate)
-100 -99.885 -100.115 Pass / Fail
-90 -89.895 -90.105 Pass / Fail
-80 -79.905 -80.095 Pass / Fail
-70 -69.915 -70.085 Pass / Fail
-60 -59.925 -60.075 Pass / Fail
-50 -49.935 -50.065 Pass / Fail
-40 -39.945 -40.055 Pass / Fail
-30 -29.955 -30.045 Pass / Fail
-20 -19.965 -20.035 Pass / Fail
-10 -9.975 -10.025 Pass / Fail
0 0.015 -0.015 Pass / Fail
10 10.025 9.975 Pass / Fail
20 20.035 19.965 Pass / Fail
30 30.045 29.955 Pass / Fail
40 40.055 39.945 Pass / Fail
50 50.065 49.935 Pass / Fail
60 60.075 59.925 Pass / Fail
70 70.085 69.915 Pass / Fail
80 80.095 79.905 Pass / Fail
90 90.105 89.895 Pass / Fail
100 100.115 99.885 Pass / Fail
Note: The maximum and minimum values listed do not take into account test equipment uncertainty, values are provided for
convenience only. Only a high performance measuring device should be used to minimize the additional uncertainty.