Page 7.6
SECTION 7 - MAINTENANCE INFORMATION
pickering
ISOLATED MILLIVOLT THERMOCOUPLE SIMULATOR 41-761
Table 7.3 - Verication Test Results For +/-50mV Range (Accuracy: 0.1% ±10μV)
Serial Number:
Channel Number:
Date of Test:
Operators Name:
Test Equipment
Used:
Voltage (mV) Measured Value Maximum (mV) Minimum (mV)
Result (delete
as appropriate)
-50 -49.940 -50.060 Pass / Fail
-45 -44.945 -45.055 Pass / Fail
-40 -39.950 -40.050 Pass / Fail
-35 -34.955 -35.045 Pass / Fail
-30 -29.960 -30.040 Pass / Fail
-25 -24.965 -25.035 Pass / Fail
-20 -19.970 -20.030 Pass / Fail
-15 -14.975 -15.025 Pass / Fail
-10 -9.980 -10.020 Pass / Fail
-5 -4.985 -5.015 Pass / Fail
0 0.010 -0.010 Pass / Fail
5 5.015 4.985 Pass / Fail
10 10.020 9.980 Pass / Fail
15 15.025 14.975 Pass / Fail
20 20.030 19.970 Pass / Fail
25 25.035 24.965 Pass / Fail
30 30.040 29.960 Pass / Fail
35 35.045 34.955 Pass / Fail
40 40.050 39.950 Pass / Fail
45 45.055 44.945 Pass / Fail
50 50.060 49.940 Pass / Fail
Note: The maximum and minimum values listed do not take into account test equipment uncertainty, values are provided for
convenience only. Only a high performance measuring device should be used to minimize the additional uncertainty.