CLEANING AND DISINFECTION OF THE APPLICATOR
After every treatment clean and disinfect the Applicator according to the following procedure:
1.
Use a soft cloth, moistened with detergent solution, to clean the Applicator until it is visually clean.
2. Use a soft cloth, moistened with water, to remove detergent residuals.
3. Use a soft, dry cloth to dry the Applicator.
4. Use a soft cloth, moistened with 70% alcohol, to disinfect the Applicator for at least 1 minute.
5. Use a soft, dry cloth to dry the Applicator.