Limitations: product is not suitable for use in the environments with a high concentration of oxygen and/or
flammable liquids and/or flammable gas; do not use with equipment for electro surgery or short-wave or
microwave therapy; the device may be interfered by other equipment, even if that other equipment complies
with CISPR EMISSION requirements.
PowerDot PD-01 has been tested to the compliance with the following Emission and Immunity standards:
Emission:
STANDARD ITEM REMARKS
CISPR 11: 2011 Conducted Class B
Radiated Class B
IEC 61000-3-2:2014 Harmonic current emissions
IEC 61000-3-3:2013 Voltage fluctuations & flicker
IMMUNITY:
STANDARD ITEM IEC 60601-1-2 Test
Levels for Home
Healthcare
Environment
PowerDot PD-01
Test Levels
REMARKS
IEC 61000-4-2:2008 ESD ± 8 kV contact;
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
±2 kV, ± 4 kV, ± 8
kV contact;
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
No
performance
degradation
observed.
IEC 61000-4-3:2010 RS 10 V/m
80 MHz – 2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz – 5.785
GHz
80% AM at 1 kHz
No
performance
degradation
observed.
IEC 61000-4-4:2012 EFT ± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition
frequency
No
performance
degradation
observed.
IEC 61000-4-5:2014 Surge ± 0.5 kV, ± 1 kV ± 0.5 kV, ± 1 kV No
performance
degradation
observed.
IEC 61000-4-6:2013 CS 3V 0.15 MHz – 80 MHz
6V in ISM and amateur
bands between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
10V 0.15 MHz –
80 MHz
80% AM at 1 kHz
No
performance
degradation
observed.
IEC 61000-4-8:2009 PFMF 30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz
No
performance
degradation
observed.
IEC 61000-4-11:2004 Voltage
dips &
voltage
variation
s
Voltage Dips:
1) 0% U
T
; 0,5 cycle at 0º.
45º, 90º, 135º, 225º, 270º,
315º
2) 0% U
T
; 1 cycle;
Single phase at 0º
3) 70% U
T
; 25/30 cycles;
Single phase at 0º
As on the previous
column
Voltage Dips:
1) No
performance
degradation
observed
2) No
performance
degradation
observed