32
Like any wireless device, PowerDot PD-02 emits very low levels in the radio frequency (RF) interval, and, is
therefore not likely to cause any interference with nearby electronic equipment (e.g. radios, computers,
telephones, etc.).
PowerDot PD-02 is designed to withstand foreseeable disturbances originating from electrostatic discharges,
mains supply magnetic fields, or radio frequency transmitters.
Despite this, it is not possible to guarantee that the stimulator will not be affected by strong RF (radio
frequency) fields emitting from other sources (such as in the proximity of working microwave oven).
Try not to use PowerDot closer than 1.5 meters to the working microwave oven as
radio interference from microwave is likely to cause disconnection between PowerDot
and your mobile phone.
Troubleshooting Wireless Connectivity
If you run into issues with Bluetooth wireless connectivity (e.g. your PowerDot PD-02 device becomes
unresponsive to PowerDot Doctor App commands during stimulation session or you were not able to connect
to your PowerDot and initiate stimulation), do not panic and consider terminating your stimulation session
manually by shortly pressing Power button on your PowerDot device.
PowerDot Doctor App has built-in re-connection and disconnection detection mechanism and, in most cases,
it will re-connect back to your PowerDot shortly and allow you to resume your stimulation using Resume
button on the stimulation screen.
If you fail to re-connect and resume stimulation after several attempts, consider stopping your stimulation
session using Stop button on the Stimulation Screen and postponing it for later.