EasyManua.ls Logo

Precise Instrument DAT 400 - Page 51

Precise Instrument DAT 400
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 49
22 0 Cod_Firmware BYTE See table C Page 45
23 0 Cod_Versione BYTE See table C Page 45
24 0 SP_1 DINT
28 0 SP_2 DINT
32 0 Load_Cell_Capacity DINT
36 0 Load_Cell_Sensitivity INT
38 0 Weight_Division_Value INT
40 0 Tare_of_the_System DINT
44 0 Capacity_of_the_System DINT
48 0 Filter_Value INT
50 0 Weight_Stability_value INT
52 0 Autozero_Set DINT
56 0 Tracking_Factor INT
58 0 Zero_Band INT
60 0 Not used BYTE
61 0 Set_1_Net_Weight BOOL See table F Page 46
61 1 Set_1_Gross_Weight BOOL See table F Page 46
61 2 Set_1_NO_NC BOOL See table F Page 46
61 3 Set_1_Pos_Neg_Values BOOL See table F Page 46
61 4 Set_1_Control BOOL See table F Page 46
62 0 Set_1_Hysteresys DINT
66 0 Set_1_Timer INT
68 0 Set_1_Delay INT
70 0 Not used BYTE
71 0 Set_2_Net_Weight BOOL See table F Page 46
71 1 Set_2_Gross_Weight BOOL See table F Page 46
71 2 Set_2_NO_NC BOOL See table F Page 46
71 3 Set_2_Pos_Neg_Values BOOL See table F Page 46
71 4 Set_2_Control BOOL See table F Page 46
72 0 Set_2_Hysteresys DINT
76 0 Set_2_Timer INT
78 0 Set_2_Delay INT
80 0 Monitor_Register INT
READING EXAMPLE
To read the gross weight on the DAT 400 it is needed to read the addresses from 2 to 5 of the Input Area.
To read the net weightit is needed to read the addresses from 6 to 9 of the Input Area.
Whwn the display shows the gross weight value of 12351 in the corresponding bytes there will be:
Byte 2 Byte 3 Byte 4 Byte 5
Hex 00 00 30 3F