Pressure Systems, Inc. Model 9116 User’s Manual
Page 38 www.PressureSystems.com
Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration
This sub-command has no additional parameters.
Command
C 03
‘C’ is the command letter.
‘ 03’ is the sub-command index (ii) for Abort.
NOTE: all parameters are separated by a space.
Response
A
‘A’ is the acknowledge letter
Description: Aborts the Multi-Point Calibration function, if it is currently in progress. This
sub-command also restores the original “A/D samples for averaging” count to the
module that was in use before the calibration function was started.
It should be noted that executing the Configure & Start (‘00’) sub-command
again, after the calibration function has started collecting data (per Collect Data
(‘01’) sub-commands), but before the final data are calculated (per Calculate &
Apply (‘02’) sub-command), will have the same affect as this Abort function.
Example:
Abort the Multi-Point Calibration function previously started
C 03
Read response:
“A”