Pressure Systems, Inc. Model 9116 User’s Manual
Page 82 www.PressureSystems.com
Example:
Send TCP/IP command to module (i.e., via its connected socket): with
replacement values for the channel’s offset and gain correction terms loaded into
the module’s volatile memory (cc=00-01). Load these into channel # 8's
Transducer Coefficient array (aa=08).
v00800-01 0.000 1.000
Response returned is:
A
Send command to Model 9116 module (via its connected socket) to change its
default EU output from psi to kPa. This will be done by changing the EU
Pressure Conversion Scaler to 6.894757.
v01101 6.894757
Response returned is
A