EasyManuals Logo

Preventice Technologies BodyGuardian One User Manual

Preventice Technologies BodyGuardian One
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
BodyGuardian OneInstructions for Use 18
Appendix B: Technical Specifications
Characteristics
Test Conditions
Min.
Typical
Max.
Unit
Physical
Length
101
mm
Width
66
mm
Thickness
13
mm
Weight
With battery
90
gm
Functional
ECG Channels
Cable selectable
2
2
3
n/a
Accelerometer
Sampling Rate
±2g/±4g/±8g
Dynamically
selectable
Variable
1.56
8
30
12
800
14
Hz
Bits
Resolution
Memory
Recording Time
30
days
Data Retention
microSDHC flash
media
10
years
Wireless
Communication Technology
4G LTE
Cellular Bands
LTE FDD 4, 13
Operating Frequencies
777
1710
787
1745
MHz
Output Power
1720.0 1745.0 MHz
0.380
W
1712.5 1752.5 MHz
0.257
782.0 782.0 MHz
0.129
779.5 784.5 MHz
0.101
Electrical
CMRR
90
105
dB
AC Range
±6
mV
DC Range
±300
mV
Input Impedance
>10
MOhm
Frequency Response
0.05 to 40
Hz
Recovery Time
1.33
sec
ADC Resolution
12
12
16
Bits
ADC Sample Rate
125
250
1000
Hz
Battery
Type
Replicable, externally
re-chargeable Li-Ion
3.7
V
Life
From full charge
72
hours
Storage Temperature
< 30 days
-20
50
C
(shipped condition)
< 30 days
-20
35
C
Relative Humidity
45
85
%
Environmental
Operating Temperature
0
40
C
Storage Temperature
-25
70
C
Relative Humidity
Non-condensing 23 C
10
93
%
Atmospheric Pressure
700
1060
hPa
Preventice Technologies reserves all rights to improve, change and modify the products and the contents of the
Instructions for Use without prior notice.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Preventice Technologies BodyGuardian One and is the answer not in the manual?

Preventice Technologies BodyGuardian One Specifications

General IconGeneral
BrandPreventice Technologies
ModelBodyGuardian One
CategoryMedical Equipment
LanguageEnglish