3302F Mainframe Operation Manual 47
CURR
Command Syntax:
[ PRESet] CCCURR{SP}{ NR2}{NL}
[ PRESet] CCCURR?{NL}
Purpose: Set and read the current.
Description: This command is used for setting of load current.
1. The least significant number is the 5
th
digit after the decimal point.
2. Should a value be entered that is higher than what is possible then the
3302F will automatically set its maximum value according the load
module fitted.
3. The engineering unit is A.
CP
{HIGHLOW}
Command Syntax:
[ PRESet] CP{ HIGHLOW}{SP}{ NR2}{NL}
[ PRESet] CP{ HIGHLOW} ? {NL}
Purpose: Set and read the operating power value in Watts
Description: This command is used for setting the HIGH and LOW setting
Levels of load power. These 2 power levels need to be used should a dynamic
Load waveform be desired. It also allow the user to switch between 2 preset
power levels.
1. The least significant number is the 5
th
digit after the decimal point.
2. The LOW level power value cannot be higher than the HIGH level.
3. Should a value be entered that is higher than what is possible then
The 3302F will automatically set its maximum value according the load
module fitted.
4. The engineering unit is W.
CRRES
{HIGHLOW}
Command Syntax:
[ PRESet] CRRES{ HIGHLOW}{SP}{ NR2}{NL}
[ PRESet] CRRES{ HIGHLOW} {NL}
Purpose: Set and read the HIGHLOW resistance levels.
Description: This command is used for setting the HIGH and LOW levels of
Load resistance. It allows the user to switch between 2 resistance levels.
1. The least significant number is the 5
th
digit after the decimal point.
2. The LOW level resistance value cannot be higher than the HIGH level.
3. Should a value be entered that is higher than what is possible then
The 3302F will automatically set its maximum value according the load
module fitted.
4. The engineering unit is Ω.