110 PRODIGIT
5.10.1.8. Press START/STOP test key.
5.10.1.9. The UUT’s output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OCP trip point is between I_Hi and I_Lo limitation, then
Middle 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
5.10.2 Remote control OCP
EX:
REMOTE
(Set Remote)
TCONFIG OCP
(Set OCP test)
OCP:START 3
(Set start load current 3A)
OCP:STEP 1
(Set step load current 1A)
OCP:STOP 5
(Set stop load current 5A)
VTH 0.6
(Set OCP VTH 0.6V)
IL 0
(Set current low limit 0A)
IH 5
(Set current high limit 5A)
NGENABLE ON
(Set NG Enable ON)
START
(Start OCP testing)
TESTING?
(Ask Testing? 1:Testing,0:Testing End)
NG?
(Ask PASS/FAIL?,0:PASS,1:FAIL)
OCP?
(Ask OCP current value)
STOP
(Stop OCP testing)