82 PRODIGIT
Measure command: available for all module
COMMAND RETURN
MEASure:CURRent{?}{;NL}
###.####
MEASure:VOLTage{?}{;NL}
###.####
MEASure:POWer{?}{;NL}
###.####
Table 4-6B MEASURE COMMAND SUMMARY
REMARK :
1. Current engineering unit: A
2. Voltage engineering unit: V
3. Period engineering unit: mS
4. Slew-rate engineering unit: (m)A/uS
5. Power engineering unit: W
Auto sequence: avilable for all module
AUTO SEQUENCE COMMAND NOTE RETURN
FILE {SP} {n}{; NL}
n=1~9 1~9
STEP {SP} {n} {;NL}
n=1~16 1~16
TOTSTEP {SP} {n}{;NL}
Total step n=1~16 1~16
SB {SP} {m,n} {;NL}
m=1~10 n=1~15
m:STATE , n:BANK
T1 {SP} {NR2} {;NL}
0.1~9.9(s) 0.1~9.9(sec)
T2 {SP} {NR2} {;NL}
0.0~9.9(s) 0.0~9.9(sec)
SAVE {;NL}
Save “File n” data
REPEAT {SP} {n} {;NL}
n=0~9999 0~9999
RUN {SP} {F} {n} {;NL}
n=1~9
AUTO REPLY
“PASS” or “FAIL:XX”
(XX=NG STEP)
Table 4-8B Auto sequence command list