EasyManua.ls Logo

Protek Z9216 - Page 14

Protek Z9216
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C+D Accuracy
The basic impedance accuracy depicted in Figure 2-1 applies to capacitance measurements when
the impedance is interpreted to be 1/2π f C, where f is the test frequency in Hz and C is the
capacitance in Farads. For convenience, Figure 2-1 is redrawn in Figure 2-3 with lines of constant
capacitance superimposed. Also, Table 2-3 is recreated for capacitive impedances as Table 2-5.
Note from Table 2-5 that the range error factor K
l
is negligible for capacitances below 1590/f µF and
K
h
is negligible for capacitances above 159/f µF.
The accuracy of the capacitance measurement is calculated from equation (1) above, with the
additional stipulation that if the measured D has a value less than 0.1, then the basic capacitance
accuracy factor should be multiplied by the factor (1 + D). Let the basic capacitance accuracy factor
be denoted A
c
.
A
c (measured)
% = ± [ Ac
× K
i
× K
v
+ 100 × (K
h
+ K
l
)] (4a)
Then the accuracy of the D calculation is given by
Accuracy of D = ± [A
c
/100] (4b)
Note that the accuracy of D is specified as a magnitude, not as a percentage.
0.55% 0.35% 0.55%
0.35% 0.25% 0.35%
0.20%
Figure 2-3 Basic Impedance Accuracy for Capacitances
1
0

Related product manuals