Table 4-1 - QDR Limits and Extreme Values
Measurement Mode QDR Limit Extreme Value
R + Q Q maximum 9999.9
L + Q Q minimum 0
C + D D maximum 9999.9
C + R series R maximum 99999
C + R parallel R minimum 0
Setting Up the Bins
Procedures
Initial Setup
To enter binning information the unit cannot be in the AUTO parameter mode. Make certain that the
unit is set to the correct measurement mode (R+Q, L+Q, C+D or C+R). Press the BIN# key, which will
bring up the bin entry display and put the unit in the entry mode. If any previous binning information
needs to be cleared, press the BIN# key until the “bin CLEAR” message appears.
Press the ENTER key to clear all bin data and display “CLEAR donE”.
To enter new bin data or to edit old bin data (if it was not previously cleared), press the BIN# key
until displays “bin x”. Enter the desired bin number (press a numeric key, 0 through 8, then ENTER).
The bin number will appear on the right alphanumeric display. This is the bin for which subsequent
nominal and limit values will be entered.
Nominal Values for Pass Bins 0 to 7
To enter the nominal value, press the NOM key. The NOM LED below the alphanumeric display will
turn on. The display will show the present nominal value, or “----“ if the bin was previously shown
closed in the left display. If a nominal value is needed for this bin, enter the desired nominal value
with the numeric keys and the unit entry keys. The new nominal value, including units, will be
displayed. Note that a nominal value does not have to be entered for each bin. If a bin does not have
a nominal value, it will use the one from the bin below it.
Limits for Pass Bins 0 to 7
To enter the limits, press the LIM key. The display will show the present upper limit value, or “----“ if
no limit was given previously. The +LIM LED below the display will turn on. Enter the limit value in
percent using the numeric keys and the ENTER key. If it is necessary to enter a non-symmetrical limit
pair, press the LIM key a second time to display the present lower limit value. The -LIM LED will turn
on. Enter the lower limit in the same fashion as the upper one. For symmetric limits, enter only the
upper value; the lower limit will be the negative of the upper limit. If no limits are entered for a bin, that
bin will remain closed, even if it has a nominal value.
4
0