Press simultaneously and hold for two seconds buttons “LEFT” (10) and 
”MODE” (12) or RC buttons (23) and (24) to enter the auxiliary menu. 
Press buttons “LEFT” (10) or “RIGHT”(11) to select menu item      
 “Display calibration”. Press “MODE” to confirm. 
Press and hold down the “MODE” (12) button for two seconds to enter the 
menu.   
Press briefly navigation buttons - “LEFT” (10) and “RIGHT” (11) to select 
icon       , press “MODE”.
Select icon       in the pop-up submenu and press “MODE”.
If you wish to return to the default defective pixel pattern (i.e. restore all 
defective pixels previously repaired), select icon in the pop-up submenu and 
press the “MODE” button.  
Options “Yes” and “No” appear on the right of the “MODE” button.
Use navigation buttons to select “Yes” and press the “MODE” button.
If you choose not to return to default pixel pattern, select “No” and press 
“MODE” button.  
To exit the main menu, press and hold down the “MODE” button for two 
seconds or wait 10 seconds to exit automatically. 
Attention! One or two pixels in the form of bright white or black 1-2 pixels 
dots are allowed on the display of thermal imager. These pixels cannot be 
repaired and are not a defect.   
A red cross appears in the centre of display, coordinates (X; Y)        of the 
cross relative to the centre of display appear in the place of pop-up icons, 
icons disappear. 
Use navigation buttons to align the center of the cross with a defective pixel 
(defective pixel should go out). Switch direction of the cross from horizontal 
to vertical by a short press of the “MODE” button.  
After the  centre  of  the  cross is  aligned  with  a defective  pixel,  press  the 
“ON/OFF” (9) button to repair the pixel. 
In case of success a short “OK” message appears next to the coordinates. 
Further on, move the cross to repair another defective pixel. When moving 
the cross to the coordinates area, the latter goes to the lower right portion of 
the display. 
To  exit  menu  option  “Defective  pixel  repair”,  press  and  hold  button 
“MODE” for two seconds. 
Return to default defective pixel pattern
X=50
Y=50
14
MAINTENANCE AND STORAGE
12
The  unit  in  monocular  mode  features  degree  of  protection  IPX7  (fully 
waterproof, submersible at 1 meter for 30 minutes); in attachment mode 
IPX5 - protection against water jets. 
Attempts to disassemble or repair the scope will void the warranty! 
Clean the scope's optical surfaces only if necessary, and use caution. First, 
remove (by blowing with a blower brush or canned air) any dust or sand 
particles. Then proceed to clean by using camera/lens cleaning equipment 
approved for use with multicoated lenses. Do not pour the solution directly 
onto the lens!
Always store the unit in its carrying case in a dry, well-ventilated space. For 
prolonged storage, remove the batteries. 
Batteries shall not be exposed to excessive heat such as sunshine, fire or the 
like. 
15
13
TROUBLESHOOTING    
Listed below are some potential problems that may occur when using the 
scope. Carry out the recommended checks and troubleshooting steps in the 
order  listed.  Please  note  that  the  table  does  not  list  all  of  the  possible 
problems. If the problem experienced with the scope is not listed, or if the 
suggested action meant to correct it does not resolve the problem, please 
contact the manufacturer. 
Problem
Possible cause Corrective action
The unit will not turn on.
The  image  is  blurry,  with 
vertical stripes and uneven 
background.
Batteries have been 
incorrectly installed.
Oxidized contact points in the 
battery compartment or 
on the battery cover due to 
“leaky” batteries or contact
 points becoming exposed 
to a chemically reactive solution.
The batteries are fully 
exhausted or one or 
several batteries are faulty.
Calibration is to be done.
Reinstall the batteries 
observing polarity.
Clean the contacts of the battery
compartment or the battery cover.
Install fresh batteries. 
Do the calibration according 
to section 9 “OPERATION”
The image is too dark. 
The lowest brightness level 
is set.
Adjust display brightness. 
Point of impact does not 
match the aiming point. 
Display calibration needs to be
done.
Do display calibration as per 
guidelines below. 
Battery cover is not tightly closed. Close tightly the battery cover.
Display calibration
Auxiliary cross in a limiting frame           and symbols will appear on the 
display       .  
First, match the auxiliary cross and the centre of sight’s reticle.
To move the frame with the cross horizontally and vertically, press buttons 
“LEFT” or “RIGHT” (or RC buttons (22),(24)). Movement takes places 
upon  both  short  and  long  press.  To  switch  movement  direction  from 
horizontal to vertical, press button “MODE” or (23) RC button. 
Second, hold “MODE” (or (23) RC button) for two seconds. Figure “2” 
and X, Y coordinates values appear next to the icon            .                             
1