QBC STAR™ System User’s Manual
5–6 L–000867(A)
External liquid controls
QBC Controls are available for additional performance monitoring of the QBC STAR system. You
and document the results before you begin testing with a new lot or newly received shipment of QBC STAR tubes. You
run liquid controls and document the results with each instance of instrument relocation or repair. Consult the
package insert accompanying the controls for preparation instructions and expected results. You
quality control requirements from your regulatory or accreditation agencies.
Profi ciency Tests
Profi ciency testing is an external evaluation of the quality of a laboratory’s performance. Laboratories enrolled in a profi ciency
System will receive fi ve unknown specimens, three times each year. These specimens are run
in the same way that patient specimens are tested. Results are submitted to the profi ciency testing program for comparison
to results obtained by other laboratories using the
A partial list of organizations that may offer profi ciency testing for the
American Profi ciency Institute (API) American College of Phycisians (ACPS)
1159 Business Park Drive 2011 Pennsylvania Ave., NW
American Academy of Family Physicians College of American Pathologists (CAP)
PT Program Coordinator Surveys Department
11400 Tomahawk Creek Parkway 325 Waukegan Road
Leawood, KS 66211 Northfi eld, Illinois 60093
800-274-7911 800-323-4040
American Association of Bioanalysts (AAB)
HCT (%)
HGB (g/dL)
MCHC (g/dL)
PLT (x 109/L)
WBC (x 109/L)
GRAN (x 109/L)
%GRANL
L/M (x 109/L)
%L/M