35
Test Procedures: 
Model 3800 
 
1. Inspection 
a.  Carefully inspect the amplifier module for burned or damaged components. 
2. Hook-up 
a.  Connect load resistors, oscilloscope channel A (20V/Div.), and distortion analyzer (1000 
Watt/0.1% input/distortion) to the amplifier output. Note: the output may be measured before 
the relay at the center tap (pins 7 and 8 of the 8-pin connector) 
b.  Connect the analyzer’s distortion output to channel B of the oscilloscope (0.1V/Div.) 
c.  Set the analyzer signal output to 2kHz, 1.25V RMS. Make all signal level changes with the 
amplifier gain controls after this adjustment is completed. 
d.  Connect the Variac (set to 0) to the amplifier’s AC inlet receptacle. 
 
Note:   Do not connect the telephone mod signal connector at this point. 
The following instructions are for 120 V models. For 240 V models, expect half the AC current draw 
at twice the AC voltage with respect to the AC measurements given. 
 
3. Initial start-up 
a.  Disconnect the load resistors from the amplifier. 
b.  Turn the amplifier on and slowly increase the Variac setting to 120VAC. Watch the 5 amp 
current meter for excessive current draw (more than 0.4 amps). The power/protect LED should 
come on red at about 20VAC. At approximately 100VAC the speaker mute relay should turn on 
and the power/protect LED should turn green. 
c.  Install the input mod connector. Observe the output for symmetrical waveform; the distortion 
analyzer should indicate less than 0.05% distortion. The signal level on the oscilloscope should 
read 160V
P-P
 (400W) when set in accordance with the above procedure (step 2c.). This verifies 
correct overall channel gain. Characteristic step spikes should be visible on the positive and 
negative halves of the waveform (still into no load) on channel B of the oscilloscope. 
d.  Switch the load to 8 ohms and vary the amplifier gain just above and below the step switching 
level. Watch for obvious step oscillation. This test must be done when the amplifier channel is 
cold since any tendency to instability will disappear as the amplifier warms up. You may 
experience a severe glitch on the negative step, which can be corrected by barely turning up 
the crossover trimmer. Verify that positive and negative peaks go into step switching within 3 
volts peak of each other. 
e.  Move the wiring harness and tap the circuit boards with a non-conductive implement such as 
the handle of a small screwdriver and watch the scope output and distortion trace for 
intermittency. Remove the 8-ohm load. 
4. DC protect 
a.  Check that there is no load on the output of the amplifier. Use a jumper to short pin 8 (+) of the 
IC to ground; the relay must turn off and the power/protect LED must turn red immediately 
(within 0.1 second). The amp should return to normal operation 2 to 3 seconds after removing 
the short. 
b.  Short pin 4 (-) of the IC to ground; the relay must turn off and the power/protect LED must turn 
red immediately (within 0.1 second). The amp should return to normal operation 2 to 3 seconds 
after removing the short. 
5. Signal indicators 
a.  With a 2kHz input signal, adjust the amp gain to obtain 50W of output power. The two yellow 
LEDs should be on at this point; the –6dB LED may be dimly lit. 
b.  Attenuate the signal by 10dB. The –6dB LED should be off. 
c.  Attenuate the signal by 20dB. The –6dB LED should be off and the –30dB LED should be dimly 
lit. 
6.  4 ohm power output and frequency response 
a.  Switch the distortion analyzer to volts/power. Select a 2kHz signal input to the amplifier. 
b.  Adjust the input signal to obtain 300 watts (x2) output and observe power output at 20Hz (-
0.5dB max.), and 20kHz (-0.5dB max.). 
7. Current limit adjustment