CONTINUOUS SAMPLING WITHOUT USING
THE INTERNAL PUMP
This analyzer provides a real-time readout of
the concentration, whether the pump button is
pressed or not. However, for continuous
measurement of a sample stream, the sample
must be at some positive pressure to force a
flow through the analyzer.
In other words, just setting the instrument in a
chamber is not enough to get a good reading.
There will never be enough gas diffusing into
the openings to cause the reading to change.
Controlling the flow rate entering the
instrument is critical when not using the pump.
We recommend 800 1000 cc/min. Lower
flow rates will produce a higher reading, and
higher flow rates will produce a lower reading.
For example, the chart below denotes the
reading when flowing 1000ppm and 100ppm
through the unit at the flow rates on the top
column:
FLOW RATE (per minute)
1.5L
1L
500cc
100cc
1000
990
1000
1028
1050
100
98
100
110
As you can see, the ideal flow rate is 1L/min
entering the analyzer.