EasyManua.ls Logo

Quantek ZR 1000 - Page 8

Quantek ZR 1000
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
If injecting your own gas, we recommend that
you use a flow meter at the inlet of the
instrument. If you measure the flow rate
leaving the instrument, it will be less, as the
internal plumbing and sensor creates a
restraint on the flow.
If you’d like to operate at a higher (or lower)
flow rate than 1000cc/min, then you will have
to recalibrate the O
2
span.
This will also require calibrating the O
2
zero.
In short, if you want to use the instrument out
of the box, then try to use a flow rate as close
to 1000 cc/min.
CONTINUOUS SAMPLING USING THE
OPTIONAL INTERNAL PUMP
The internal sampling pump (optional) will draw
about 5cc per second (~300cc/min) from your
sample gas, flow, or chamber.
When using the pump, it is recommended that
you have sufficient sample for analysis. In
other words, if the instrument takes 5 minutes
for a full response, and you use the internal
pump, it will draw a total of 1.5L of sample.
This makes sampling a 2L chamber
impossible, unless you return the sample back
to the chamber.