42 www.quantumrehab.com Quantum 6000
FLASHING LIGHTS DIAGNOSIS SOLUTION
10 High Battery Voltage Check batteries.
9 Solenoid Brake Fault Check motor/brake wiring.
8 Possible Controller Fault See Quantum Rehab Specialist.
7 Possible Joystick Fault See Quantum Rehab Specialist.
6 Inhibit Active Unplug charger. Check connections.
5 Right Motor Wiring Fault Check right motor wiring.
4 Right Motor Disconnected Check right motor wiring.
3 Left Motor Wiring Fault Check left motor wiring.
2 Left Motor Disconnected Check left motor wiring.
1 Low Battery Voltage Check batteries/battery wiring.