Description Reactor-Ready
Jacket Temperature Range -60˚C to +190˚C
Hose Fittings M24
Weight 27kg
Stirrer Speed
500rpm continuous
8
00rpm for short periods
10. Routine Maintenance
10.1 To maintain your Reactor-Ready system in prime condition it is important that care is
taken to check all moving parts and seals regularly.
10.2 Moving parts should be kept clean and free from any obstructions or surface
contamination. Parts relying on sliding motion may benefit from regular application of a thin
surface coating of lubricating oil.
10.3 The 3 coupling pins on both the Stirrer Guide and Stirrer Drive Coupling should be
regularly inspected. These pins will wear over time and will require periodic replacement.
10.4 The Quick Release Insulated Hose features a double o-ring sealing mechanism within
the PEEK coupling. To ensure that this coupling provides a good seal, regular inspection of
the o-rings is recommended. O-rings should be replaced if any sign of damage or wear is
observed, or if any leakage occurs from this connection during operation.
10.5 The FEP encapsulated silicone o-ring Seal in the PTFE O-ring Support Collar will also
require periodic replacement.
10.6 Reactor-Ready metal components are resistant to the majority of solvents and splash
resistant to dilute acids and alkalis at room temperature. However, always wipe away any
chemical spills immediately and avoid exposure to strong vapours such as HCL etc, as
continued exposure to even low levels of chemical vapours or liquids will cause corrosion to
metal and/or plastic components. Elevated temoeratures will also accelerate any potential
corrosion/chemical attack.
Page 17
9. General Specifications
Reactor-Ready - Lab Reactor