- 91 -
Response options:
RV_A_”x” CR LF
- command understood, response: program version
RV_I CR LF
- command understood but not accessible at this moment
xprogram version (in between inverted commas)
An example:
Command:
RV CR LF
return program version
Response:
RV_A_” 1.1.1”
program version: ”1.1.1”
Set AUTOZERO function
Format: A_n CR LF
Response options:
A_OK CR LF
command carried out
A_E CR LF
error in-course of command execution, no parameter or incorrect format
A_I CR LF
– command understood but not accessible at this moment
n parameter, decimal value determining autozero settings
n → 0 autozero off
1 autozero on
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
A_1 CR LF
turn autozero function on
Response:
A_OK CR LF
autozero function is on
AUTOZERO function operates until it is turned off by A 0 command.
Set ambient conditions state
Format: EV_n CR LF
Response options:
EV_OK CR LF
command carried out
EV_E CR LF
error in-course of command execution, no parameter or incorrect format
EV_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining ambient conditions state
n 0 unstable ambient conditions
1 stable ambient conditions
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
EV_1 CR LF
set value ‘stable’ for ambient conditions option
Response:
EV_OK CR LF
ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps it to
value <UNSTABLE>.
Set filter
Format: FIS_n CR LF
Response options:
FIS_OK CR LF
command carried out
FIS_E CR LF
error in-course of command execution, no parameter or incorrect format
FIS_I CR LF
command understood but not accessible at this moment
n parameter, decimal value determining filter number
n → 1 very fast
2 fast
3 average
4 slow
5 very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.