- 92 -
Set Value release
– error in-course of command execution, no parameter or incorrect format
– command understood but not accessible at this moment
n – parameter, decimal value determining value release options
2 – fast+reliable
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release
settings are assigned to the working mode
– set value release parameter to fast+reliable option
– fast+reliable option set
Response options:
– error in-course of command execution, no parameter or incorrect format
– command understood but not accessible at this moment
n – parameter, decimal value determining last digit settings
2 – never
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit settings
are assigned to the working mode.
– set last digit option to value ‘always’
Cooperation with PUE 7.1, PUE 10 terminal
- command not understood (format error)
- response: mass value in a basic measuring unit
1
2
3
4
5
6
7
8
9-18
19
20-22
23
24-32
33
34-36
37
38
39
40
N
T
space
stability
marker
zero marker
range
marker
digit marker
space
mass
space
mass unit
space
tare
space
tare unit
space
digits
CR
LF
- [space] if stable, [?] if unstable
- [space] for any value but zero, [Z] for zero value
- marker informing on range within which mass value is comprised:
[space] I range balance, [2] II balance, [3] III balance