Ambient conditions module – parameter specifying module’s
working mode. Available options: None – re
condition module indications turned off; Record and alert –
readout and record of indications into database turned on, option
of display of warnings informing on ambient conditions change
turned on, the warnings are displayed in accordanc
following tolerance values settings.
Additional temperature sensor –
indication of the additional temperature sensor connected to the
ambient conditions module.
Min temperature –
min balance temperature, for lower
temperature values the thermometer pictogram is red
Max temperature –
max balance temperature, for higher
temperature values the thermometer pictogram is red
Temperature ∆t/h –
maximum balance temperature change
rate, for higher change rate values the balance displays blinking
red thermometer pictogram
Min humidity – min humidity value, for lower humidity values
the thermometer pictogram is red
Max humidity – max humidity value, for higher humidity values
the thermometer pictogram is red
Humidity ∆%/h – maximum hum
idity change rate, for higher
change rate values the balance displays blinking red
thermometer pictogram
With all the parameters set, the home screen displays respective pictograms informing on current
ambient conditions readouts and their change.
Pictograms for ambient conditions state:
Indicated temperature is within specified permissible limits
2
Indicated temperature is out of specified permissible limits
3
Indicated humidity is within specified permissible limits
4
Indicated humidity is out of specified permissible limits
5
Temperature change rate is to high (blinking pictogram)
6
Humidity change rate is too high (blinking pictogram)