- 127 -
Set ambient conditions state
Format: EV_n <CR><LF>
Response options:
EV_OK <CR><LF>
command carried out
EV_E <CR><LF>
error in-course of command execution, no parameter or incorrect format
EV_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining ambient conditions state
n → 0 unstable ambient conditions
1 stable ambient conditions
CAUTION!
Command changes settings for a current working mode.
An example:
Command:
EV_1<CR><LF>
set value ‘stable’ for ambient conditions option
Response:
EV_OK<CR><LF>
ambient conditions option set to value ‘stable’
<AMBIENT CONDITIONS> parameter is set to value <STABLE> until command EV 0 swaps it to
value <UNSTABLE>.
Set filter
Format: FIS_n <CR><LF>
Response options:
FIS_OK <CR><LF>
command carried out
FIS_E <CR><LF>
error in-course of command execution, no parameter or incorrect format
FIS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining filter number
n → 1 very fast
2 fast
3 average
4 slow
5 very slow
CAUTION!
The numbering is assigned to a particular filter name and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, filter settings are
assigned to the working mode.
An example:
Command:
FIS_3<CR><LF>
set average filter
Response:
FIS_OK<CR><LF>
average filter set
Set Value release
Format: ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
command carried out
ARS_E <CR><LF>
error in-course of command execution, no parameter or incorrect format
ARS_I <CR><LF>
command understood but not accessible at this moment
n parameter, decimal value determining value release options
n → 1 fast
2 fast+reliable
3 reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value release
settings are assigned to the working mode
An example:
Command:
ARS_2<CR><LF>
set value release parameter to fast+reliable option
Response:
ARS_OK<CR><LF>
fast+reliable option set