EasyManuals Logo

RADWAG PS X2 Series User Manual

RADWAG PS X2 Series
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
- 49 -
Time
Time parameter enables setting the current time. Procedures for change of time settings and date
settings are likewise.
Date format
Date form. parameter enables altering the date format on the printout [YYYY.MM.DD / YYYY.DD.MM
/ DD.MM.YYYY / MM.DD.YYYY], where: YYYY year; MM month; DD day.
Time format
Time form. parameter enables specifying time format for a printout [12h / 24h].
For [12h] option selected, <A> or <P> letter is displayed next to presented time value, where: A
stands for hours before noon; P stands for hours after noon.
Backlight turn-off time
<BACKLIGHT OFF> parameter enables activation of display stand-by mode, the stand-by mode is
activated when no weighing process is carried out (stable indication is a necessary condition for
activation of the stand-by mode).
NONE backlit turn-off time not activated.
0.5; 1; 2; 3; 5time given in minutes.
If the software registers stable indication for a specified time interval, set in parameter <BACKLIGHT
OFF>, than the display goes out immediately. The backlight activates upon change of indication (no
stability pictogram on the display) or pressing any key on the balance keypad. The display remains
blank also when balance menu is entered.
Auto switch-off
<AUTO OFF> parameter enables automatic display deactivation (the parameter functioning is
likewise to
button functioning). Upon display deactivation the other subassemblies are powered
and the balance turns to stand-by mode.
NONE auto switch-off not activated.
0.5; 1; 2; 3; 5time given in minutes.
If the software registers stable indication for a specified time interval, set in parameter <AUTO OFF>,
than the display is turned-off immediately.
To start-up the balance, it is necessary to press
button located on the balance keypad. The
balance automatically returns to weighing operation.
Balance cannot be turned off if any process is started or if balance menu is entered.
Autotest GLP
AUTOTEST GLP function is designed to aid a user in assessing balance’s operation and diagnosing
the reasons for occurrence of errors in weighing which exceed the maximum permissible values for a
given balance model.
By means of a simple, repeatable and fully documented way, the function enables optimizing
balance’s settings to maintain the best possible repeatability and weighing time at workstation. The
main purpose of the function is the possibility of monitoring the above-mentioned parameters at

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS X2 Series and is the answer not in the manual?

RADWAG PS X2 Series Specifications

General IconGeneral
BrandRADWAG
ModelPS X2 Series
CategoryScales
LanguageEnglish

Related product manuals