EasyManua.ls Logo

R&S NRPxxS Series - Trace Measurement with Synchronization to Measurement Complete

R&S NRPxxS Series
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performing Measurement Tasks - Programming Examples
R&S
®
NRPxxS(N)
149User Manual 1177.5079.02 ─ 10
//Query the measurement results
FETCh?
10.5 Trace Measurement with Synchronization to Measure-
ment Complete
This example, written in pseudo code, shows how to setup and execute a trace mea-
surement using a non-blocking technique.
The advantage of using the FETCH? command (as shown in the previous example) is,
that FETCH? will wait (block) until a measurement result is available. However this may
lead to situations where an application blocks for a longer time (until timeout) if, for
example, a trigger is missing and thus no results will ever become available.
For certain applications, especially interactive ones, it is not the desired behavior that
the user has to wait until a (probably long) timeout occurs. In these cases it is preferred
to start a measurement and then enter a loop to poll the sensor until the measurement
is ready and the results can safely be retrieved. For such applications it is recommen-
ded to use the status system of the sensor to find out whether the measurement is
ready. The advantage of this approach is that the polling loop can be exited/canceled
at any time and the application stays operable (i. e. does not block).
// basic setup, similar to the previous example
write( "*RST" );
write( "SENS:FUNC \"XTIM:POW\"" );
write( "SENS:FREQ 1.8e9" );
write( "SENS:TRAC:POIN 500" );
write( "SENS:TRAC:TIME 20e-3" );
write( "TRIG:SOUR INT" );
write( "TRIG:SLOP POS" );
write( "TRIG:DTIM 0.001" );
write( "TRIG:HYST 0.1" );
write( "TRIG:LEV 30e-6" );
write( "SENS:TRAC:AVER:COUN 8" );
write( "SENS:TRAC:AVER:STAT ON" );
// configuring the event system to recognize the
// end of measurement (i.e. a negative transition
// of bit 1 in the meas operation register)
write( "STAT:OPER:MEAS:NTR 2" );
write( "STAT:OPER:MEAS:PTR 0" );
// resetting the event information by an initial readout
int iEvent = 0;
query( "STAT:OPER:MEAS:EVEN?", &iEvent );
// Now starting the measurement
write( "INIT:IMM" );