EasyManua.ls Logo

R&S NRPxxS Series - Controlling the Measurement

R&S NRPxxS Series
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
NRPxxS(N)
72User Manual 1177.5079.02 ─ 10
See also Chapter 9.7.1, "Continuous Average Measurement", on page 93.
Burst average ("POWer:BURSt:AVG"): The integration time of a measurement is
not predefined but determined by the sensor with the aid of a burst detector. The
start of a burst is detected when the measurement signal rises above a set trigger
level. The measurement ends when the signal drops below a trigger threshold.
See also Chapter 9.7.2, "Burst Average Measurement", on page 96.
Timeslot average ("POWer:TSLot:AVG"): The power is simultaneously measured
in up to 32 time slots. The measurement result is represented by a vector that can
contain up to 32 indices and contains the power of a time slot at each index.
See also Chapter 9.7.3, "Timeslot Average Measurement", on page 97.
Trace ("XTIMe:POWer"): A sequence of measurements is performed.
See also Chapter 9.7.4, "Trace Measurement", on page 99.
9.5 Controlling the Measurement
The power sensor offers a bunch of possibilities to control the measurement:
Do you want to start the measurement immediately after the initiate command or
do you want to wait for a trigger event?
Do you want to start a single measurement cycle or a sequence of measurement
cycles?
Do you want to output each new average value as a measurement result or do you
want to bundle more measured values into one result?
9.5.1 Starting and Ending a Measurement
ABORt............................................................................................................................ 72
INITiate:ALL.................................................................................................................... 73
INITiate[:IMMediate]......................................................................................................... 73
INITiate:CONTinuous........................................................................................................73
ABORt
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (INITiate[:IMMediate]), the sensor goes into the idle
state. However, if a continuous measurement is in progress (INITiate:CONTinuous
ON), the trigger system of the sensor enters the waiting for trigger state, and if the trig-
ger condition is met, a new measurement is immediately started.
See also Chapter 9.5, "Controlling the Measurement", on page 72.
Usage:
Event