Operating concepts
R&S
®
NRT-Z14/44
23Getting Started 1443.8632.02 ─ 01
5 Operating concepts
● Base unit......................................................................................................... 23
● R&S Virtual NRT............................................................................................. 24
● Remote control................................................................................................26
5.1 Base unit
See also Chapter 3.5.2, "Base unit", on page 17.
Setup
2 3
1
4
5
Figure 5-1: Setup for base unit
1 = signal source
2 = R&S NRT‑Zxx directional power sensor
3 = load
4 = base unit
5 = host interface connector → sensor interface
Communication between the R&S NRT‑Zxx directional power sensor and a base
unit is only possible with a baud rate setting of 38400 Bd. This setting is the fac-
tory default that must be restored if the setting was changed. If the R&S NRT‑Zxx
directional power sensor is not recognized by the base unit, check that the baud
rate setting of the R&S NRT‑Zxx directional power sensor is 38400 Bd.
For details on checking and setting the baud rate, see the user manual > mainte-
nance.
Base unit