Remote Control
R&S
®
RTO
634User Manual 1316.0827.02 ─ 06
Parameters:
<Stop> Range:  -100E+24  to  100E+24
Increment:  0.01
*RST:  0.01
Parameters for setting and query: 
<SearchName> Search definition
Usage:  SCPI confirmed
SEARch:GATE:RELative:STARt <SearchName>,<RelativeStart>
SEARch:GATE:RELative:STARt? <SearchName>
Defines the starting value for the gate.
Parameters:
<RelativeStart> Range:  -100E+24  to  100E+24
Increment:  0.1
*RST:  0
Default unit: %
Parameters for setting and query: 
<SearchName> Search definition
Usage:  SCPI confirmed
SEARch:GATE:RELative:STOP <SearchName>,<RelativeStop>
SEARch:GATE:RELative:STOP? <SearchName>
Defines the end value for the gate.
Parameters:
<RelativeStop> Range:  -100E+24  to  100E+24
Increment:  0.1
*RST:  100
Default unit: %
Parameters for setting and query: 
<SearchName> Search definition
Usage:  SCPI confirmed
SEARch:GATE:ZCOupling <SearchName>,<ZoomCoupling>
SEARch:GATE:ZCOupling? <SearchName>
If enabled, the gate area is defined identically to the zoom area for the zoom diagram.
Parameters:
<ZoomCoupling> ON | OFF
*RST:  OFF
Parameters for setting and query: 
<SearchName> Search definition
Command Reference