Command reference
R&S
®
ZNA
1028User Manual 1178.6462.02 ─ 20
Parameters:
<Boolean> ON | OFF - Limit check on or off. A result is available even if the
limit check is disabled; see example for CALCulate<Chn>:
RIPPle:SEGMent<Seg>:RESult?.
*RST: n/a (no ripple limit line defined after a reset)
Example: See CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
STARt
Manual operation: See "Range List" on page 443
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit
<Limit>
Defines the ripple limit for ripple limit range no. <Seg>. A range must be created first to
enable this command (e.g. CALCulate<Chn>:RIPPle:DATA). See Chapter 4.4.1.2,
"Rules for ripple test definition", on page 144.
Tip: To define several ripple limit ranges with a single command, use
CALCulate<Chn>:RIPPle:DATA.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Number of the ripple limit range.
Parameters:
<Limit> Ripple limit in the range. The unit is adjusted to the format of the
active trace (CALCulate<Chn>:FORMat).
Range: Virtually no restriction for ripple limit ranges.
*RST: n/a (no ripple limit line defined after a reset)
Default unit: See above.
Example: See CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:
STARt
Manual operation: See "Range List" on page 443
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
Returns the result of the ripple limit check in the previously defined limit range no.
<Seg>. The response consists of two parameters:
<Boolean> 0 for "passed", 1 for "failed".
<Limit> measured ripple in the limit range. A result is returned even if the limit
check in the range no. <Seg> is disabled; see example below.
A reset deletes all ripple limit ranges. Use CALCulate<Ch>:RIPPle:FAIL? to query
the result for global ripple limit check.