Command reference
R&S
®
ZNA
1238User Manual 1178.6462.02 ─ 20
Parameters:
<Factor> Slope factor
*RST: 0
Default unit: dB/GHz
Options: R&S ZNA-K8
Manual operation: See "Slope" on page 833
[SENSe:]CONVerter<Port>:IFPort <IfInput>
Selects input connectors to be used for the IF meas and IF ref connection between the
converter and the analyzer port <Port>.
Suffix:
<Port>
.
Converter number = VNA port number
Parameters:
<IfInput> DRACcess | REAR
DRACcess
Use the (optional) direct access connectors at the front panel of
the R&S ZNA.
REAR
Use the (optional) direct IF access connectors IF Reference <i>
and IF Meas <i> at the rear panel of the R&S ZNA.
Options: R&S ZNA-K8
Manual operation: See "IF Input" on page 829
[SENSe:]CONVerter<Port>:IFRequency <IF>
Sets/queries the intermediate frequency for converter <Port>.
Suffix:
<Port>
.
Converter number = VNA port number
Parameters:
<IF> Intermediate frequency
Defaults to the intermediate frequency ([SENSe:]CONVerter:
DEFinition:IFRequency on page 1234) of the connected
converter type ([SENSe:]FREQuency:CONVersion:
DEVice<Port>:NAME).
Default unit: Hz
Options: R&S ZNA-K8
Manual operation: See "IF Frequency" on page 829
[SENSe:]CONVerter<Port>:LOLoss <Loss>
Defines the cable loss at port <Port> if used as LO port (without splitter).
SCPI command reference