Command reference
R&S
®
ZNA
1319User Manual 1178.6462.02 ─ 20
<SourceType> PORT | GENerator
PORT - internal source at port <CalPort>, to be fed to port
<SourcePort> using an external through connection. If
<CalPort> = <SourcePort>, an Open or Short standard is
required.
<SourcePort> Number of the port for the internal source or generator.
<CalOnlyPortFreq> Relevant for frequency conversion measurements only.
OFF | 0
The reference receiver at port <CalPort> is calibrated for all fre-
quencies that are relevant for any of the ports.
ON | 1
Only those frequencies are calibrated that are relevant for this
port. This results in shorter calibration times.
*RST: 0
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is 1).
CALC:PAR:MEAS 'TRC1', 'A1'
Select a
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no. <Mk> in the center of the sweep range and
query the measurement value. The calibrated power of the refer-
ence wave a
1
is approx. equal to the default source power value.
See also example for [SENSe<Ch>:]CORRection:
OFFSet<PhyPt>[:STATe].
Usage:
Setting only
Manual operation: See "Port Overview" on page 543
[SENSe<Ch>:]CORRection:POWer<PhyPt>:AWAVe[:STATe] <Boolean>
Enables or disables the receiver power calibration for channel <Ch> and for the refer-
ence waves a
n
. The setting command is disabled unless the reference waves have
been power calibrated ([SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
on page 1318 AWAVe,...). The query always returns a result.
This command extends the functionality of manual control, where the reference
receiver calibration is included in the source power calibration of each port (and always
ON).
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number