Command reference
R&S
®
ZNA
1320User Manual 1178.6462.02 ─ 20
Parameters:
<Boolean> Enables (ON) or disables (OFF) the receiver power calibration
for the reference waves a
n
.
*RST: OFF
Example: See [SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire
Starts the receiver calibration
at physical port <PhyPt>
in channel <Ch>
at harmonic orders [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer?
?
Then stores and applies the calibration data.
Note: The receiver calibration relies on the source power calibration acquired using
SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire].
Suffix:
<Ch>
.
Channel number
<PhyPt> Port number.
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
HARMonic[:ACQuire] on page 1443.
Usage: Event
Manual operation: See "Harmonic Orders to Calibrate" on page 536
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire
[<CalOnlyPortFreq>]
Starts the receiver calibration (3
rd
power calibration step for intermodulation measure-
ments), stores and applies the calibration data.
Note: The receiver calibration relies on the source power calibration acquired in step
nos. 1 and 2.
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire is enabled
only after steps 1 and 2 have been carried out (SOURce<Ch>:POWer<PhyPt>:
CORRection:IMODulation:LTONe[:ACQuire], SOURce<Ch>:POWer<PhyPt>:
CORRection:IMODulation:UTONe[:ACQuire]).
Suffix:
<Ch>
.
Calibrated channel number
SCPI command reference