Command reference
R&S
®
ZNA
1343User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]FREQuency:STARt <StartFreq>
[SENSe<Ch>:]FREQuency:STOP <StopFreq>
These commands defines the start and stop frequency for a frequency sweep. The val-
ues also define the display range in a Cartesian diagram. The default start and stop
frequencies equal to the minimum and maximum frequency of the analyzer.
The ranges depend on the instrument model; see Table 7-17.
Suffix:
<Ch>
.
Channel number
Parameters:
<StopFreq> Start and stop frequency of the sweep
Increment: 0.1 kHz
Default unit: Hz
Example:
*RST; FREQ:STAR 100000
Activate a frequency sweep and set the start frequency to 100
kHz.
FREQ:STOP 10MHz
Set the stop frequency to 10 MHz.
Manual operation: See "Start Frequency / Stop Frequency / Center Frequency /
Span Frequency" on page 474
Note: If the start frequency entered is greater than the current stop frequency, the stop
frequency is set to the start frequency plus the minimum frequency span
([SENSe<Ch>:]FREQuency:SPAN).
If the stop frequency entered is smaller than the current start frequency, the start fre-
quency is set to the stop frequency minus the minimum frequency span.
7.3.14.13 [SENSe:]GDEVice
Configures generic device commands for the related channel.
SENSe:GDEVice:ADDRess?.........................................................................................1343
[SENSe<Ch>:]GDEVice:DELete....................................................................................1344
[SENSe<Ch>:]GDEVice:MSELect..................................................................................1344
[SENSe<Ch>:]GDEVice:SELect.................................................................................... 1344
SENSe:GDEVice:ADDRess?
Returns the addresses of the configured generic external devices.
Return values:
<Address> Comma-separated list of string values.
These addresses are used for setting and unsetting the com-
mand to be executed via [SENSe<Ch>:]GDEVice:SELect
and [SENSe<Ch>:]GDEVice:DELete, respectively.
Example: See [SENSe<Ch>:]GDEVice:SELect on page 1344.
Usage: Query only
SCPI command reference