Command reference
R&S
®
ZNA
1360User Manual 1178.6462.02 ─ 20
<Pt> Port number
Usage: Query only
Options: R&S ZNA-B26
Manual operation: See "Rear IF Frequency" on page 623
[SENSe<Ch>:]PATH<Pt>:MEASurement:IFSWitch?
[SENSe<Ch>:]PATH<Pt>:REFerence:IFSWitch?
Queries if and how the Direct IF Access Input / Output ports IF Reference <Pt> and IF
Meas <Pt> are used in channel <Ch>.
To define the port usage, use [SENSe<Ch>:]PATH<Pt>:IFSWitch.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Usage: Query only
Options: R&S ZNA-B26
Manual operation: See "Rear Out" on page 622
7.3.14.19 [SENSe:]PHASe...
[SENSe<Ch>:]PHASe:MODE <True Phase Mode>
Defines the phase mode of channel <Ch>, i.e. de/activates coherence mode in combi-
nation with low phase noise mode.
Suffix:
<Ch>
.
Channel number
Parameters:
<True Phase Mode> NCOHerent | COHerent | LNNCoherent | LNCoherent
NCOHerent
Coherence off, low phase noise off
COHerent
Coherence on, low phase noise off
LNNCoherent
Coherence off, low phase noise on
LNCoherent
Coherence on, low phase noise on
*RST: NCOHerent
Manual operation: See "Phase Mode" on page 634
SCPI command reference