10 Thermalert GP Series Operator’s Manual
2.4.1 Display/Key Pad
Before turning on the power, make sure all wiring connections are secure.
Table 5 shows the factory-set default values.
Note that all controls can be adjusted while the power is on without damag-
ing sensor or electronics.
Table 5: Factory-set Default Values
Emissivity
Sensing Head/Device
Number
Display Resolution
Setpoints (SP1, SP2)
Peak Hold
Valley Hold
Averaging
Ambient Temperature
Compensation (t-amb)
Offset
0.10 - 1.09
in 0.01 increments
H001 - H006
0 - 4
According to sensing
head/device type
0 - 998 seconds
999 = infinite hold
0 - 998 seconds
999 = infinite hold
0 - 60 seconds
-18 to 1200°C
(0 to 2200°F)
-50 to 50°C
(-99 to 99°F)
For 0 - 5V and 4 - 20 mA &
thermocouple input,
is not functional
See Section 2.4.2
See Sections 2.4.2 and 2.4.4
Infinite reset by hardware
Infinite reset by hardware
0.95
H001
0
off
0
0
0
0
0
PARAMETERS SETTING RANGE REMARKS
DEFAULT
VALUES