11
Ambient temperature offset
1. Hold the SET button and then press the HOLD button to increase
the value.
2. Hold the SET button and then press the BKLT button to decrease
the value.
3. Press the MODE and POWER buttons to save the value.
4. TheLCDdisplaywillindicate"SAVE"conrmingtheselection.
5. Repower the meter to resume normal operation.
Humidity Offset
1. Hold the MAX/MIN button and then press the HOLD button to
increase the value.
2. Hold the MAX/MIN button and then press the BKLT button to
decrease the value.
3. Press the MODE and POWER buttons to save the value.
4. TheLCDdisplaywillindicate"SAVE"conrmingtheselection.
5. Repower the meter to resume normal operation.
Pressure Offset Function
1. While in normal operation, press and hold the SET button until the
LCD shows "0.0 hpa".
2. To increase the value press the
HOLD button.
3. To decrease the value press the
BKLT button.
4. Press and hold the POWER button
to save the set value.
5. TheLCDdisplaywillindicate"SAVE"conrmingtheselection.
Note: The compensation range is ±3 hpa.
Error Codes
Code Description
Er1: Sensor failed.
Er2: Out of measured range.
WBGT WBGT
WBGT
GT
%
hpa
AVG
MIN
MAX
WBGT
HIDPT
%
hpa
GT
%
%
REC
hpa
3
4
5
continued...
1.888.610.7664 info@REED-Direct.com
REED-Direct.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Reed Instruments R6200 and is the answer not in the manual?

Reed Instruments R6200 Specifications

General IconGeneral
BrandReed Instruments
ModelR6200
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals