7
www.REEDINSTRUMENTS.com
3. WhentheLCDashes,pressthe REL button to enter calibration
mode.Thiswillbeindicatedby"CAL1"ontheLCD.
4.
Insertthetipofthemeterintothe33%RHHumidityCalibrationStandard
(REEDR9933)thenpresstheMAX/MIN button to start the adjustment.
The calibration adjustment will be conrmed by a long beep.
5.
Themeterwillthendisplay"CAL2"tocalibratethesecondcalibration
point.Insertthetipofthemeterintothe75%RHHumidityCalibration
Standard(REEDR9975)thenpressMAX/MIN button to start the adjust-
ment. The calibration adjustment will be conrmed by a long beep.
6. After both calibrations are complete, the meter will resume normal
operation.
Note: The calibration steps can be cancelled at any point by pressing the
ON/OFFbutton.
Restoring Factory Settings
1. Turn the meter off.
2.
PresstheHOLD,°C/°F,andpowerbuttonsimultaneously,thenrelease.
3. WhentheLCDashes,presstheMAX/MIN button to restore factory
default settings.
4. A long beep will conrm factory settings have been restored. The
meter will then resume normal operation.
Battery Replacement
When the low battery indicator appears on the display, the battery should be
replaced.Removethebatterycoveronthebackandinsertanew9Vbattery.
Accessories and Replacement Parts
CA-05A Soft Carrying Case
R993333%HumidityCalibrationStandard
R997575%HumidityCalibrationStandard
R998033%and75%HumidityCalibrationKit