7
www.REEDINSTRUMENTS.com
3. WhentheLCDashes,pressthe REL button to enter calibration
mode.Thiswillbeindicatedby"CAL1"ontheLCD.
4.
Insertthetipofthemeterintothe33%RHHumidityCalibrationStandard
(REEDR9933)thenpresstheMAX/MIN button to start the adjustment.
The calibration adjustment will be conrmed by a long beep.
5.
Themeterwillthendisplay"CAL2"tocalibratethesecondcalibration
point.Insertthetipofthemeterintothe75%RHHumidityCalibration
Standard(REEDR9975)thenpressMAX/MIN button to start the adjust-
ment. The calibration adjustment will be conrmed by a long beep.
6. After both calibrations are complete, the meter will resume normal
operation.
Note: The calibration steps can be cancelled at any point by pressing the
ON/OFFbutton.
Restoring Factory Settings
1. Turn the meter off.
2.
PresstheHOLD,°C/°F,andpowerbuttonsimultaneously,thenrelease.
3. WhentheLCDashes,presstheMAX/MIN button to restore factory
default settings.
4. A long beep will conrm factory settings have been restored. The
meter will then resume normal operation.
Battery Replacement
When the low battery indicator appears on the display, the battery should be
replaced.Removethebatterycoveronthebackandinsertanew9Vbattery.
Accessories and Replacement Parts
CA-05A Soft Carrying Case
R993333%HumidityCalibrationStandard
R997575%HumidityCalibrationStandard
R998033%and75%HumidityCalibrationKit
Don't see your part listed here? For a complete list of all accessories and
replacement parts visit your product page on www.reedinstruments.com.