6. Error Messages/Troubleshooting
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(Example Error no. 1)
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined
but after wards the cuff pressure fell
below 20 mmHg.
This situation can occur, for example, if
the tube becomes unplugged after the
systolic blood pressure has been
measured. Further possible cause: No
pulse has been detected.
ERR 2 Unnatural pressure impulses influence
the measurement result. Reason: The
arm was moved during the
measurement (Artefact).
ERR 3 Inflation of the cuff takes too long. The
cuff is
not correctly seated or the hose
connection is not sealed.
ERR 5 The difference between systolic and
diastolic is excesive. Measure again
carefully following proper cuff
procedures and ensure measurement
under quiet conditions. Contact your
doctor if this message occurs
frequently.
10