English 1
Disinfection and sterilization guide Clinical use only
This guide is intended for multipatient use of the AirFit N20 / AirFit N20 for Her ResMed nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User
Guide for cleaning instructions. This guide describes ResMed’s validated procedures for cleaning, disinfection and sterilization of the mask in accordance with ISO17664.
AirFit N20 / AirFit N20 for Her
mask component
1
High level thermal disinfection High level chemical disinfection Sterilization Validated number of
cycles
2
Thermal water bath
194°F (90°C)
for 1 min
Thermal water bath
199.4-203°F (93-95°C)
for 10 min
CIDEX
™
OPA
Ortho-phthalaldehyde 0.55%
12 min soak
STERRAD
™
100S
Short cycle
(boosters not required)
STERRAD
™
NX
Standard or
advanced cycle
• Cushion
—
30
• Magnetic clip
— — — —
• Multi-hole elbow and short tube
3
— — — —
• Frame / headgear
— — — —
1
This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide.
2
If a healthcare facility requires an additional disinfection or sterilization cycle after reassembly, the number of validated cycles must be halved.
3
The multi-hole elbow and short tube are not separable.
A
B C D
2
6
7
1
4
5
8
3
A Multi-hole elbow and short tube
1 Elbow
2 Side buttons
3 Vent
4 Short tube
5 Swivel
B Frame
C Cushion
D Headgear
6 Magnetic clips
7 Lower headgear straps
8 Upper headgear straps
A+B+C Frame system
A+B+C+D Complete system