English 1
Disinfection and sterilisation guide Clinical use only
This guide is intended for multipatient use of the AirFit N20 / AirTouch N20 and AirFit N20 Classic ResMed nasal masks in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the
User Guide for cleaning instructions. This guide describes ResMed’s recommended and validated procedures for cleaning, disinfection and sterilisation of the mask in accordance with ISO17664.
AirFit N20 / AirTouch N20
AirFit N20 / AirTouch N20
mask component
1
Thermal disinfection Chemical disinfection Sterilisation Validated
number of
cycles
2
Manual AWD
5
CIDEX
™
OPA Gigasept FF Anioxyde 1000 STERRAD 100S STERRAD NX
• AirFit N20 cushion 30
• AirTouch N20 cushion
3
– – – – – – – –
• Magnetic clips
– – –
30
• Elbow and short tube
4
–
30
• Frame
– – –
30
• Headgear
– – –
30
1
This mask may not be available in all regions. For full details regarding the correct use of these masks, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Product Guide on ResMed.com.
2
If a healthcare facility requires an additional disinfection or sterilisation cycle after reassembly, the number of validated cycles must be halved.
3
The AirTouch N20 cushion is intended for single-patient reuse in both home and hospital/institutional environments and recommended to be replaced monthly.
4
The elbow and short tube are not separable.
5
Automatic Washer-Disinfector